6 Points Chemiluminescence Calibrator Kit For POCT Analyzer POClia With 8 Channels

Ask Latest Price
Video Channel
Brand Name :Zecen
Model Number :POCT Chemiluminescence Calibrator Kit
Certification :CE, ISO13485
Place of Origin :China
MOQ :1
Price :Negotiable
Payment Terms :L/C, D/A, D/P, T/T, Western Union, MoneyGram
Supply Ability :5000 Set/Sets per Month
Delivery Time :10 workdays
Packaging Details :Wooden case or carton package
Item :Chemiluminescence Calibrator Kit for POCT
Analyzer :POClia in Zecen
Principle :6-points
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

6-points Chemiluminescence Calibrator Kit for POCT Analyzer POClia with 8 Channels

product composition Six points: ABCDEF
Testing method 1. Treat the calibrator as the sample to be tested, and place it in the sample loading area of the instrument after mixing. Note that when mixing Avoid air bubbles.
2. During the detection, the system detects the luminescence value (RLU value) of each calibration point, and calculates the luminescence value according to the nominal concentration and luminescence value.
The calibration curve is automatically fitted by the instrument. In order to ensure the reliability of the calibration curve, the determination of the calibrator should be
Make a double.
3. The generated calibration curve can be used for 28 days, but it should be re-calibrated after the following conditions:
① Change of batch number of supporting testing reagents;
Change of batch number of luminescent substrate
②The measured value of the quality control product is not within the scope of quality control or the test results are inconsistent with the clinical situation;
③System for major repairs or maintenance.
Interpretation of test results 1. For reasons such as methodology or antibody specificity, the same sample was tested using reagents from different manufacturers.
Testing with different reagents may give different test results, therefore, the results obtained by testing with different reagents should not be directly
They are compared with each other in order to avoid false medical interpretations.
2. The quality control product can be used as a reference for the reliability of the current experimental results, and its measured value should be within the
within the range allowed by the quality control sheet. Test results should be based on reference value ranges and other clinical factors and results
Comprehensive judgment, when the test result is close to the upper limit or lower limit of the reference value range, it can be considered
samples for confirmation testing.
Storage and expiration date Stored at 2-8°C, the product is valid for 12 months. Reagents should be used within 1 month after opening, and stored at 2-8°C away from direct sunlight.