CLIA Calibrator of Inflammation
Inflammation Item | CIV | |
LN | ||
HA | ||
CG | ||
Six-point Calibrator | Calibrator A | |
Calibrator B | ||
Calibrator C | ||
Calibrator D | ||
Calibrator E | ||
Calibrator F | ||
Calibrator Performance | Appearance | Each component should be complete, the internal and external packagipg should be complete, the label should be clear, and there should be no leakage of liquid reagent. |
Volume | The volume of each component shall be accurately weighed, and it shall not be lower than the nominal volume. | |
Accuracy | The relative deviation should be within ± 10%. | |
Homogeneity | The homogeneity between bottles CV ≤ 10%. | |
Stability | Under the storage condition of 2 ℃ ~ 8 ℃, the validity period of IL-6 calibrator is 12 months. Take the samples stored within 1 month after the validity period to test the appearance, volume, accuracy and uniformity index, which should meet the standard requirements. |