
Add to Cart
Inflammation Reagent Calibrator PCT CRP IL-6 SAA Reagent Kit Six-Points Accuracy Calibtarors IVD Healthcare Diagnostic
Testing method:
1. Treat the calibrator as the sample to be tested, and place it in the sample loading area of the instrument after mixing. Note that when mixing Avoid air bubbles.
2. During the detection, the system detects the luminescence value (RLU value) of each calibration point, and calculates the luminescence value according to the nominal concentration and luminescence value.
The calibration curve is automatically fitted by the instrument. In order to ensure the reliability of the calibration curve, the determination of the calibrator should be make a double.
Storage and expiration date:
Stored at 2-8°C, the product is valid for 18 months.
Reagents should be used within 1 month after opening, and stored at 2-8°C away from direct sunlight.
CLIA/POCT Calibrator of Inflammation
Inflammation Item | PCT | |
CRP | ||
IL-6 | ||
SAA* | ||
Six-point Calibrator | Calibrator A | |
Calibrator B | ||
Calibrator C | ||
Calibrator D | ||
Calibrator E | ||
Calibrator F | ||
Calibrator Performance | Appearance | Each component should be complete, the internal and external packagipg should be complete, the label should be clear, and there should be no leakage of liquid reagent. |
Volume | The volume of each component shall be accurately weighed, and it shall not be lower than the nominal volume. | |
Accuracy | The relative deviation should be within ± 10%. | |
Homogeneity | The homogeneity between bottles CV ≤ 10%. | |
Stability | Under the storage condition of 2 ℃ ~ 8 ℃, the validity period of IL-6 calibrator is 12 months. Take the samples stored within 1 month after the validity period to test the appearance, volume, accuracy and uniformity index, which should meet the standard requirements. |